Skip to main content

Scanner quick reference

For sonographers and nurses. Print this page and keep it by the scanner: the site navigation is left off the printout.

Before you rely on this card

Your clinic's scanners must already be connected and have passed a test scan (see the clinic setup guide). If they have not, scans will not reach wawa however carefully you follow it.

Every scan, every scanner​

  1. Choose the patient from the worklist. Never type their details by hand: the worklist entry carries the wawa ID that puts the scan on the right record. If someone is missing, refresh the worklist.
  2. Scan and measure as usual. For follicles, set left or right before measuring each ovary.
  3. Send the images.
  4. Send the measurements.
  5. End the exam.
ImagesMeasurements
Skipped?The scan shows in wawa with measurements but no imagesThe scan shows in wawa with images but Measurements pending

Both sends, every scan.

GE Voluson P8​

StepOn the P8
WorklistPatient key, then the Worklist tab. Choose the patient and Start Exam
Send the imagesPress P1 once all the images are captured
Send the measurementsOpen Measurements and press Transfer Data (bottom left of the window)
FinishEnd Exam

GE Voluson Performance and Swift​

These follow the same two sends as the P8, but wawa has not verified their controls, and they depend on the model, software version and how your supplier set the scanner up:

  • Images are usually sent with P1, if your supplier set it to send to the imaging server.
  • Measurements may be sent from the report screen, or automatically at End Exam if the scanner is set to.

Fill in your scanner's controls after your go-live test scan, not before:

Our scanner (model and room): ______________________
Send the images with 
Send the measurements with 

Any other scanner​

The same two sends are needed on every make and model; only the buttons differ. wawa does not publish model-specific steps beyond the P8, so ask your scanner supplier which controls send each one, and write them in the box above. For GE ViewPoint, see ViewPoint.

Check it in wawa​

Open the patient, then Records → Ultrasound. Scans arrive within a few minutes; to bring them across now, click Fetch Scanner Reports. Images can take up to an hour to show. Measurements come into wawa for follicle-tracking and pregnancy scans only; any other scan shows its images and stays Measurements pending.

Quick fixes​

What you seeMost likelyWhat to do
Patient missing from the worklistNot booked as an ultrasound, or booked for another roomRefresh the worklist; check the booking's appointment type and room. Still missing? Tell your wawa administrator
Measurements but no images, on every scan from this scannerThe image key does not send to wawa's imaging server: unset, or still pointing at an old systemNot something a sonographer can fix: tell whoever looks after the scanners to check P1 in Button Configuration
Measurements but no images, on some scansThe image send was missedRe-send the exam from the scanner's archive with P1, and tell wawa the date
Measurements showing, images not yetImages are downloaded in the background and can take up to an hourWait before re-sending. Still missing after an hour? See the rows above
Images showing, Measurements pendingThe measurement send was missedSend the measurements, then Fetch Scanner Reports. Still pending? Tell wawa
Nothing in wawa at allNothing was sent, or the patient was typed by handSend the images and the measurements, then Fetch Scanner Reports. If it was typed by hand, tell wawa

Still stuck? Tell wawa the patient's wawa ID (not their name), the scan date and time, and the scanner or room.